Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0213520060200040225
Korean Journal of Ophthalmology
2006 Volume.20 No. 4 p.225 ~ p.229
The Relationship between Optical Coherence Tomography and Scanning Laser Polarimetry Measurements in Glaucoma
Chung Yun-Suk

Sohn Yong-Ho
Abstract
Purpose: To investigate the relationship between optical coherence tomography (OCT) and scanning laser polarimetry (SLP) in measuring peripapillary retinal nerve fiber layer (RNFL) thickness in glaucomatous eyes.

Methods: Fifty glaucomatous eyes were evaluated in this study. Evaluations were analyzed two ways. First, parameters of the Stratus OCT (average thickness, superior/ inferior average) and GDx VCC (TSNIT average, nerve fiber indicator (NFI), superior/ inferior average) were correlated using the Pearson¡¯s correlation coefficient (r). Secondly, comparison (r) of these parameters was completed using the mean deviation (MD) of visual field defect.

Results: The following parameters were found to be significantly correlated (P<0.005). TSNIT average/average thickness (r=0.673), NFI/average thickness (r=-0.742), superior average (r=0.841), and inferior average (r=0.736). In the correlation analysis using the severity of visual field defect, all these parameters had statistically meaningful correlations (P<0.005).

Conclusion: GDx VCC and Stratus OCT are highly correlated in glaucomatous eyes. Therefore, peripapillary RNFL thickness measured by Stratus OCT and GDx VCC may be equally helpful in the diagnosis of glaucoma. Korean Journal of Ophthalmology 20(4):225-229, 2006
KEYWORD
GDx VCC, Glaucoma, OCT, Retinal nerve fiber layer thickness
FullTexts / Linksout information
 
Listed journal information
MEDLINE ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø