KMID : 0213520080220040242
Korean Journal of Ophthalmology
2008 Volume.22 No. 4 p.242 ~ p.245
Anterior Chamber Measurements by Pentacam and AS-OCT in Eyes With Normal Open Angles
Yi Jeong-Ho

Hong Sa-Min
Ma Kyoung-Tak
Kim Chan-Yun
Kang Sung-Yong
Seong Gong-Je
Abstract
Purpose: To assess the reproducibility and agreement of anterior chamber measurements between the Pentacam (PTC) and the Anterior segment optical coherence tomography (AOCT) in normal healthy eyes with open angle.

Methods: Prospective cross-sectional comparative case series. A total of 162 eyes of 81 healthy volunteers with normal open angle were included in this study. Anterior chamber angle (ACA) and anterior chamber depth (ACD) were measured with PTC and AOCT. Intra-observer variability and inter-methods agreement of both instruments for ACA and ACD were evaluated.

Results: Values of temporal and nasal ACA measured by two instruments were similar, and the results of ACD were also not significantly different between modalities (p>0.01). ACA and ACD measurements by PTC and AOCT showed good intra-observer and inter-method agreements (all >0.9).

Conclusions: PTC and AOCT are presumed to be very useful for the anterior chamber angle examination. They may provide good images and quantitative data about the angle structures including ACA and ACD.
KEYWORD
Anterior chamber angle, Anterior chamber depth, Anterior segment optical coherence tomogaphy, Pentacam