Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0213520230370050387
Korean Journal of Ophthalmology
2023 Volume.37 No. 5 p.387 ~ p.394
Effect and Safety of Pressure Sensor-equipped Handpiece in Phacoemulsification System
Kim Jin-Yeong

Kim Ha-Rin
Jun Ik-Hyun
Kim Tae-Im
Seo Kyoung-Yul
Abstract
Purpose: To evaluate the effect of the Active Sentry handpiece of the Centurion Vision System compared to the Centurion Ozil handpiece for phacoemulsification in cataract surgery.

Methods: A retrospective study was conducted on 281 patients (449 eyes) who underwent cataract surgery between August 2020 and June 2021. Preoperative measurements, intraoperative parameters, complication rate, and postoperative outcomes were compared between the Active Sentry handpiece and the Centurion Ozil handpiece groups. Additionally, the parameters were compared in different cataract severity groups and multiple predictive factors for the number of active surge mitigation (ASM) actuations were assessed with the Active Sentry handpiece.

Results: There were 198 eyes in the Active Sentry group and 251 eyes in the Centurion Ozil group. There were no statistically significant differences between the two groups, as the cumulative dissipated energy in the Active Sentry and Centurion Ozil groups were 8.32 ¡¾ 7.74 and 7.87 ¡¾ 9.25 ¥ìJ, respectively (p = 0.576). Total surgery time, ultrasound usage time, aspiration time, amount of fluid aspirated, postoperative corrected distant visual acuity, and postoperative decrease in corneal endothelial cell density were comparable between the two groups. The significant contributors to the number of ASM actuations were age, preoperative corrected distant visual acuity, axial length, and total ultrasound time.

Conclusions: There was no clear advantage of the Active Sentry handpiece compared to the Centurion Ozil handpiece. ASM actuation increases with age, poor visual acuity before surgery, short axial length, and prolonged ultrasound usage time. It is expected that in more severe and high-risk cataract surgery, the Active Sentry handpiece functions more effectively, possibly affecting the safety and prognosis.
KEYWORD
Active fluidics, Active sentry handpiece, Cataract, Phacoemulsification, Retrospective study
FullTexts / Linksout information
 
Listed journal information