KMID : 0311120170580061186
|
|
Yonsei Medical Journal 2017 Volume.58 No. 6 p.1186 ~ p.1194
|
|
Three-Dimensional Fast Spin-Echo Imaging without Fat Suppression of the Knee: Diagnostic Accuracy Comparison to Fat-Suppressed Imaging on 1.5T MRI
|
|
Cho Hee-Woo
Suh Jin-Suck Park Jin-Oh Kim Hyoung-Sik Chung Soo-Yoon Lee Young-Han Hahn Seok
|
|
Abstract
|
|
|
Purpose: To evaluate the diagnostic performance of three-dimensional fast spin-echo (3D FSE-Cube) without fat suppression (NFS) for detecting knee lesions, using comparison to 3D FSE-Cube with fat suppression (FS).
Materials and Methods: One hundred twenty-four patients who underwent 1.5T knee magnetic resonance imaging (MRI) scans and 25 subsequent arthroscopic surgeries were retrospectively reviewed. Using arthroscopic results and two-dimensional images as reference standards, diagnostic performances of 3D FSE-Cube-NFS and FS imaging about lesions of ligament, meniscus, subchondral bone marrow edema (BME), and cartilage were compared. Scan parameters of 3D FSE-Cube imaging were previously optimized by a porcine knee phantom.
Results: No significant differences were observed between detection rates of NFS and FS imaging for detecting lesions of meniscus and cartilage (p>0.05). However, NFS imaging had lower sensitivity for detection of medial collateral ligament (MCL) tears, and lower sensitivity and specificity for detection of BME lesions, compared to FS imaging (p<0.05).
Conclusion: 3D FSE-Cube-NFS imaging showed similar diagnostic performance for detecting lesions of meniscus or cartilage compared to FS imaging, unlike MCL or BME lesions.
|
|
KEYWORD
|
|
Knee, magnetic resonance imaging, three-dimensional image, Cube, fat suppression
|
|
FullTexts / Linksout information
|
|
|
|