KMID : 0338420110260040421
The Korean Journal of Internal Medicine
2011 Volume.26 No. 4 p.421 ~ p.426
Sound Analysis in an In Vitro Endotracheal Tube Model
Park Young-Sik

Kee Young-Wook
Park Kwang-Suk
Lee Jin-Woo
Lee Sang-Min
Yim Jae-Joon
Yoo Chul-Gyu
Kim Young-Whan
Han Sung-Koo
Yang Seok-Chul
Abstract
Background/Aims: Complete endotracheal tube obstruction is a medical emergency, and partial occlusion causes increased breathing rates and failure to wean off mechanical ventilation. Partial occlusion may be underestimated due to the lack of proper detection methods. We tested whether the sound of an endotracheal tube could be used to detect an endotracheal tube obstruction using an in vitro model.

Methods: An endotracheal tube was connected to a ventilator on one end and a test lung on the other. Sounds were recorded with a microphone located inside the endotracheal tube via a connector. During mechanical ventilation, we changed the endotracheal tube internal diameter from 5.0 to 8.0 mm and different grades of obstruction at different sites were used along the tube. Sound energy was compared among the different conditions.

Results: The energy of endotracheal tube sounds was positively correlated with the internal diameter and negatively correlated with the degree of obstruction. The rate of decline in energy differed with obstruction location. When the obstruction was more distal, the rate of decline in endotracheal sound energy was more rapid.

Conclusions: Changes in the sound of an endotracheal tube can be used to detect an obstruction. Further studies are needed for clinical application.
KEYWORD
Intubation, intratracheal, Sound, Airway obstruction