KMID : 0368120090390020071
|
|
Korean Circulation Journal 2009 Volume.39 No. 2 p.71 ~ p.74
|
|
Clinical Utility and Accuracy of Coaguchek¢ç XS, a Reliable Alternative to Laboratory International Normalized Ratio Monitoring in Korean Patients With Atrial Fibrillation
|
|
Kim Sung-Hea
Suh Soon-Yong Kim Hyun-Joong Han Seong-Woo Jung Sang-Man Ryu Kyu-Hyung
|
|
Abstract
|
|
|
Background and Objectives: To evaluate the accuracy and the clinical utility of the Coaguchek¢ç XS, a portable point-of-care coagulometer, compared to standard laboratory methods in Korean patients
Subjects and Methods: We included 107 patients with atrial fibrillation on long-term oral warfarin therapy. The prothrombin time as expressed by the international normalized ratio (INR) was measured by the portable Coaguchek¢ç XS system as well as standard laboratory methods. Agreement between the two methods was defined as a difference of less than 0.3 between the INR values.
Results: The INR determined by the portable Coaguchek¢ç XS had excellent correlation with the values obtained by standard methods (r=0.984, p<0.01). The INR determined by the portable Coaguchek¢ç XS tended to be high among the high INR group values (standard INR>3.0). There was a 91.6% agreement between the two methods. Only nine cases (8.4%) had a difference of more than 0.3; this was mainly noted in the high INR group.
Conclusion: The portable self-testing of the INR by the Coaguchek¢ç XS might be a reliable alternative to hospital based laboratory testing in Korean patients.
|
|
KEYWORD
|
|
Atrial fibrillation, Prothrombin time
|
|