Korean Circulation Journal
2016 Volume.46 No. 2 p.179 ~ p.185
Tachyarrhythmia Cycle Length in Appropriate versus Inappropriate Defibrillator Shocks in Brugada Syndrome, Early Repolarization Syndrome, or Idiopathic Ventricular Fibrillation
Lee Woo-Seok

Kim Jun
Kwon Chang-Hee
Choi Jin-Hee
Jo Uk
Kim Yoo-Ri
Nam Gi-Byoung
Choi Kee-Joon
Kim You-Ho
Abstract
Background and Objectives: Implantable cardioverter?defibrillators (ICDs) are indicated in patients with Brugada syndrome (BS), early repolarization syndrome (ERS), or idiopathic ventricular fibrillation (IVF) who are at high risk for sudden cardiac death. The optimal ICD programming for reducing inappropriate shocks in these patients remains to be determined. We investigated the difference in the mean cycle length of tachyarrhythmias that activated either appropriate or inappropriate ICD shocks in these three patient groups to determine the optimal ventricular fibrillation (VF) zone for minimizing inappropriate ICD shocks.

Subjects and Methods: We selected 41 patients (35 men) (mean age¡¾standard deviation=42.6¡¾13.0 year) who received ICD shocks between April 1996 and April 2014 to treat BS (n=24), ERS (n=9), or IVF (n=8). Clinical and ICD interrogation data were retrospectively collected and analyzed for all events with ICD shocks.

Results: Of the 244 episodes, 180 (73.8%) shocks were appropriate and 64 (26.2%) were inappropriate. The mean cycle lengths of the tachyarrhythmias that activated appropriate and inappropriate shocks were 178.9¡¾28.7 ms and 284.8¡¾24.4 ms, respectively (p<0.001). The cutoff value with the highest sensitivity and specificity for discriminating between appropriate and inappropriate shocks was 235 ms (sensitivity, 98.4%; specificity, 95.6%). When we programmed a single VF zone of ¡Â270 ms, inappropriate ICD shocks were reduced by 70.5% and appropriate shocks were missed in 1.7% of these patients.

Conclusion: Programming of a single VF zone of ¡Â270 ms in patients with BS, ERS, or IVF could reduce inappropriate ICD shocks, with a low risk of missing appropriate shocks.