Design and Filtration Performance of Polysulfone Hollow Fiber Membrane Hemofilter
|
|
±èÀçÁø
¹ÚÁø¿ë/¼»óºÀ/±è¼º¼ö/±èÀº¿µ/°¼ºÁ¾/¿¬Àμ®/°í¿µÈ¯
|
|
Abstract
|
|
|
Hemofilter was optimally designed and manufactured using polysulfone hollow fiber mem- brane with surface area of 0.6mE Molecular weight cut-off of the hemofilter was measured with polyethylene glycol and dextran aqueous solutions of various molecular weights and it was ranged from 9,500 to 38,900. Ultrafiltration rates were measured with pure water in a static system and flowing system. The clearance of urea, creatinlne, and vitamine BIB were measured using aqueous solutions.
|
|