Objective: We wanted to evaluate the impact of two reconstruction algorithms (halfscan and multisector) on the image quality and the accuracy of measuring the severity of coronary stenoses by using a pulsating cardiac phantom with different heart rates (HRs).

Materials and Method : Simulated coronary arteries with different stenotic severities (25, 50, 75%) and different luminal diameters (3, 4, 5 mm) were scanned with a fixed pitch of 0.16 and a 0.35 second gantry rotation time on a 64-slice multidetector CT. Both reconstruction algorithms (halfscan and multi-sector) were applied to HRs of 40-120 beats per minute (bpm) at 10 bpm intervals. Three experienced radiologists visually assessed the image quality and they manually measured the stenotic severity.

Results: Fewer measurement errors occurred with multisector reconstruction (p = 0.05), a slower HR (p < 0.001) and a larger luminal diameter (p = 0.014); measurement errors were not related with the observers or the stenotic severity. There was no significant difference in measurements as for the reconstruction algorithms below an HR of 70 bpm. More nonassessable segments were visualized with halfscan reconstruction (p = 0.004) and higher HRs (p < 0.001). Halfscan reconstruction had better quality scores when the HR was below 60 bpm, while multisector reconstruction had better quality scores when the HR was above 90 bpm. For the HRs between 60 and 90 bpm, both reconstruction modes had similar quality scores. With excluding the nonassessable segments, both reconstruction algorithms achieved a similar mean measured stenotic severity and similar standard deviations.

Conclusion: At a higher HR (above 90 bpm), multisector reconstruction had better temporal resolution, fewer nonassessable segments, better quality scores and better accuracy of measuring the stenotic severity in this phantom study.
KEYWORD
Computed tomography (CT), Reconstruction, Algorithm, Phantom, Coronary artery angiography