Objective: The aim of this study was to evaluate the usefulness of combined fluoroscopy- and CT-guided transthoracic needle biopsy (FC-TNB) using a cone beam CT system in comparison to fluoroscopy-guided TNB (F-TNB).
Materials and Methods: We retrospectively evaluated 74 FC-TNB cases (group A) and 97 F-TNB cases (group B) to compare their respective diagnostic accuracies according to the size and depth of the lesion, as well as complications, procedure time, and radiation dose.
Results: The sensitivity for malignancy and diagnostic accuracy for small (< 30 mm in size) and deep (¡Ã 50 mm in depth) lesions were higher in group A (91% and 94%, 92% and 94%) than in group B (73% and 81%, 84% and 88%), however not statistically significant (p > 0.05). Concerning lesions ¡Ã 30 mm in size and < 50 mm in depth, both groups displayed similar results (group A, 91% and 92%, 80% and 87%; group B, 90% and 92%, 86% and 90%). Pneumothorax occurred 26% of the time in group A and 14% for group B. The mean procedure time and patient skin dose were significantly higher in group A (13.6 ¡¾ 4.0 minutes, 157.1 ¡¾ 76.5 mGy) than in group B (9.0 ¡¾ 3.5 minutes, 21.9 ¡¾ 15.2 mGy) (p < 0.05).
Conclusion: Combined fluoroscopy- and CT-guided TNB allows the biopsy of small (< 30 mm) and deep lesions (¡Ã 50 mm) with high diagnostic accuracy and short procedure times, whereas F-TNB is still a useful method for large and superficial lesions with a low radiation dose.
|
|