Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0880420220230020246
Korean Journal of Radiology
2022 Volume.23 No. 2 p.246 ~ p.255
Usefulness of Silent MRA for Evaluation of Aneurysm after Stent-Assisted Coil Embolization
Kim You-Na

Choi Jin-Wook
Lim Yong-Cheol
Song Ji-Hye
Park Ji-Hyun
Jung Woo-Sang
Abstract
Objective: To determine the usefulness of Silent MR angiography (MRA) for evaluating intracranial aneurysms treated with stent-assisted coil embolization.

Materials and Methods: Ninety-nine patients (101 aneurysms) treated with stent-assisted coil embolization (Neuroform atlas, 71 cases; Enterprise, 17; LVIS Jr, 9; and Solitaire AB, 4 cases) underwent time-of-flight (TOF) MRA and Silent MRA in the same session using a 3T MRI system within 24 hours of embolization. Two radiologists independently interpreted both MRA images retrospectively and rated the image quality using a 5-point Likert scale. The image quality and diagnostic accuracy of the two modalities in the detection of aneurysm occlusion were further compared based on the stent design and the site of aneurysm.

Results: The average image quality scores of the Silent MRA and TOF MRA were 4.38 ¡¾ 0.83 and 2.78 ¡¾ 1.04, respectively (p < 0.001), with an almost perfect interobserver agreement. Silent MRA had a significantly higher image quality score than TOF MRA at the distal internal carotid artery (n = 57, 4.25 ¡¾ 0.91 vs. 3.05 ¡¾ 1.16, p < 0.001), middle cerebral artery (n = 21, 4.57 ¡¾ 0.75 vs. 2.19 ¡¾ 0.68, p < 0.001), anterior cerebral artery (n = 13, 4.54 ¡¾ 0.66 vs. 2.46 ¡¾ 0.66, p < 0.001), and posterior circulation artery (n = 10, 4.50 ¡¾ 0.71 vs. 2.90 ¡¾ 0.74, p = 0.013). Silent MRA had superior image quality score to TOF MRA in the stented arteries when using Neuroform atlas (4.66 ¡¾ 0.53 vs. 3.21 ¡¾ 0.84, p < 0.001), Enterprise (3.29 ¡¾ 1.59 vs. 1.59 ¡¾ 0.51, p = 0.003), LVIS Jr (4.33 ¡¾ 1.89 vs. 1.89 ¡¾ 0.78, p = 0.033), and Solitaire AB stents (4.00 ¡¾ 2.25 vs. 2.25 ¡¾ 0.96, p = 0.356). The interpretation of the status of aneurysm occlusion exhibited significantly higher sensitivity with Silent MRA than with TOF MRA when using the Neuroform Atlas stent (96.4% vs. 14.3%, respectively, p < 0.001) and LVIS Jr stent (100% vs. 20%, respectively, p = 0.046).

Conclusion: Silent MRA can be useful to evaluate aneurysms treated with stent-assisted coil embolization, regardless of the aneurysm location and type of stent used.
KEYWORD
Silent magnetic resonance angiography, Time-of-flight MRA, Stent-assisted coil embolization, Stent design, Zero TE MRA