KMID : 0893320020170010037
Journal of Environmental Toxicology
2002 Volume.17 No. 1 p.37 ~ p.51
Application and Optimization of the IsoButoxycarbonyl Derivatization method to the Analysis of Trace Level Phenols in Environmental Samples
Kim Hyub

Hong Jong-Ki
Kim Yong-Hwa
Kim Kyung-Rae
Abstract
Eleven phenols including two chlorophenols, eight alkylphenols and bisphenol A were derivatized with isobutylchloroformate to form their isobutoxycarbonyl derivatives. Standard phenol mixture was concentrated for the isobutoxycarbonyl (isoBOC) derivatization and analysed by gas chromatography/mass spectrometry (GC/MS). The recoveries of the derivatization method of alkylphenols, chlorophenols, and bisphenol A were calculated by gas chromatography/mass spectrometry-selected ion monitoring mode using two work-up methods for comparison; shaking and heating method. The linear detector responses were obtained in the concentration range of 5¡­400 ng, with correlation coefficients varying from 0.9755¡­0.9981. Recoveries of the alkylphenols, chlorophenols, and bisphenol A were determined by gas chromatography/mass spectrometry-selected ion monitoring mode using two work-up methods for comparison ; the US-EPA method and the isoBOC derivatization method, Eleven phenols in water samples were extracted with dichloromethane and then concentrated. Also, solid-phase extraction (SPE) with XAD-4 and subsequent conversion to isobutoxycarbonyl derivatives for sensitive analysis with the selected ion-monitoring (SIM) mode. The recoveries were 85.1¡­109.9£¥ and 90.3¡­126.6£¥ for the US-EPA method and the isoBOC