Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1023820180100050354
The Journal of Advanced Prosthodontics
2018 Volume.10 No. 5 p.354 ~ p.360
Accuracy of provisional crowns made using stereolithography apparatus and subtractive technique
Kang Seen-Young

Park Jung-Hyun
Kim Ji-Hwan
Kim Woong-Chul
Abstract
PURPOSE: To compare and analyze trueness and precision of provisional crowns made using stereolithography apparatus and subtractive technology.

MATERIALS AND METHODS: Digital impressions were made using a master model and an intraoral scanner and the crowns were designed with CAD software; in total, 22 crowns were produced. After superimposing CAD design data and scan data using a 3D program, quantitative and qualitative data were obtained for analysis of trueness and precision. Statistical analysis was performed using normality test combined with Levene test for equal variance analysis and independent sample t-test. Type 1 error was set at 0.05.

RESULTS: Trueness for the outer and inner surfaces of the SLA crown (SLAC) were 49.6¡¾9.3 ¥ìm and 22.5¡¾5.1 ¥ìm, respectively, and those of the subtractive crown (SUBC) were 31.8¡¾7.5 ¥ìm and 14.6¡¾1.2 ¥ìm, respectively. Precision values for the outer and inner surfaces of the SLAC were 18.7¡¾6.2 ¥ìm and 26.9¡¾8.5 ¥ìm, and those of the SUBC were 25.4¡¾3.1 ¥ìm and 13.8¡¾0.6 ¥ìm, respectively. Trueness values for the outer and inner surfaces of the SLAC and SUBC showed statistically significant differences (P<.001). Precision for the inner surface showed significance (P<.03), whereas that for the outer surface showed no significance (P<.58).

CONCLUSION: The study demonstrates that provisional crowns produced by subtractive technology are superior to crowns fabricated by stereolithography in terms of accuracy.
KEYWORD
Stereolithography apparatus, Subtractive technology, Provisional crown, CAD/CAM system
FullTexts / Linksout information
 
Listed journal information
SCI(E) ÇмúÁøÈïÀç´Ü(KCI) KoreaMed