Progress in Medical Physics
2013 Volume.24 No. 1 p.35 ~ p.40
Evaluation of Absorbed Dose and Skin Dose with MDCT Using Ionization Chamber and TLD
Jeon Kyung-Soo

Oh Young-Kee
Baek Jong-Geun
Kim Ok-Bae
Kim Jin-Hee
Choi Tae-Jin
Jeong Dong-Hyeok
Kim Jeung-Kee
Abstract
Recently, the uses of Multi-Detector Computed Tomography (MDCT) for radiation treatment simulation and planning which is used for intensity modulated radiation therapy with high technique are increasing. Because of the increasing uses of MDCT, additional doses are also increasing. The objective of this study is to evaluate the absorbed dose of body and skin undergoing in MDCT scans. In this study, the exposed dose at the surface and the center of the cylindrical water phantom was measured using an pencil ionization chamber, 30 cc ionization chamber and TL Powder. The results of MDCT were 31.84 mGy, 33.58 mGy and 32.73 mGy respectively. The absorbed dose at the surface showed that the TL reading value was 33.92 mGy from MDCT. These results showed that the surface dose was about 3.5% from the MDCT exposure higher than a dose which is located at the center of the phantom. These results mean that the total exposed dose undergoing MDCT 4 times (diagnostic, radiation therapy planning, follow-up et al.), is about 14 cGy, and have to be considered significantly to reduce the exposed dose from CT scan.
KEYWORD
MDCT , CTDI , Pencil ion chamber , 30 cc ion-chamber , TLD