Progress in Medical Physics
2013 Volume.24 No. 2 p.127 ~ p.132
Evaluation of the Accuracy for Respiratory-gated RapidArc
Sung Ji-won

Yoon Myong-Geun
Chung Weon-Kyu
Bae Sun-Hyun
Shin Dong-Oh
Kim Dong-Wook
Abstract
The position of the internal organs can change continually and periodically inside the body due to the respiration. To reduce the respiration induced uncertainty of dose localization, one can use a respiratory gated radiotherapy where a radiation beam is exposed during the specific time of period. The main disadvantage of this method is that it usually requests a long treatment time, the massive effort during the treatment and the limitation of the patient selection. In this sense, the combination of the real-time position management (RPM) system and the volumetric intensity modulated radiotherapy (RapidArc) is promising since it provides a short treatment time compared with the conventional respiratory gated treatments. In this study, we evaluated the accuracy of the respiratory gated RapidArc treatment. Total sic patient cases were used for this study and each case was planned by RapidArc technique using varian ECLIPSE v8.6 planning machine. For the Quality Assurance (QA), a MatriXX detector and I¡¯mRT software were used. The results show that more than 97% of area gives the gamma value less than one with 3% dose and 3 mm distance to agreement condition, which indicates the measured dose is well matched with the treatment plan¡¯s dose distribution for the gated RapidArc treatment cases.
KEYWORD
Gated RapidArc, RPM system, Respiratory gated radiotherapy, Treatment planning system, Dose distribution
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed