KMID : 1038720130240030191
Progress in Medical Physics
2013 Volume.24 No. 3 p.191 ~ p.197
The Development of Real Time Automatic Patient Position Correction System during the Radiation Therapy Based on CCD: A Feasibility Study
Shin Dong-Ho

Chung Kwang-Zoo
Kim Me-young
Son Jae-Man
Yoon Myong-Geun
Lim Young-Kyung
Lee Se-Byeong
Abstract
Upon radiation treatment, it is the important factor to monitor the patient¡¯s motion during radiation irradiated, since it can determine whether the treatment is successful. Thus, we have developed the system in which the patient¡¯s motion is monitored in real time and moving treatment position can be automatically corrected during radiation irradiation. We have developed the patient¡¯s position monitoring system in which the patient¡¯s position is three dimensionally identified by using two CCD cameras which are orthogonal located around the isocenter. This system uses the image pattern matching technique using a normalized cross-correlation method. We have developed the system in which trigger signal for beam on and off is generated by quantitatively analyzing the changes in a treatment position through delivery of the images taken from CCD cameras to the computer and the motor of moving couch can be controlled. This system was able to automatically correct a patient¡¯s position with the resolution of 0.5 mm or less.
KEYWORD
3D moving couch, Position correction, Patient positioning, Gating