KMID : 1038720130240030198
Progress in Medical Physics
2013 Volume.24 No. 3 p.198 ~ p.203
Quantitative Evaluation of Gated Radiation Therapy Using Gamma Index Analysis
Ma Sun-Young

Choi Ji-Hoon
Jeung Tae-Sig
Lim Sang-Wook
Abstract
Generally, to evaluate gated radiation therapy, moving phantoms are used to simulate organ motion. Since the target moves in every direction, we need to take into account motion in each direction. This study proposes methods to evaluate gated radiation therapy using gamma index analysis and to visualize adequate gating window sizes according to motion ranges. The moving phantom was fabricated to simulate motion in the craniocaudal direction. This phantom consisted of a moving platform, the I¡¯m MatriXX, and solid water phantoms. A 6 MV photon filed with a field size of 4¡¿4 cm2 was delivered to the phantom using the gating system, while the phantom moved in the 1-, 2-, 3-, 4-, and 5-cm motion ranges. The gating windows were set at 40¡­60%, 30¡­40%, and 0¡­90%, respectively. The I¡¯m MatriXX acquired the dose distributions for each scenario and the dose distributions were compared with a 4¡¿4 cm2 static filed. The tolerance of the gamma index was set at 3%/3 mm. The greater the gating window, the lower the pass rate, and the greater the motion range, the lower the pass rate in this study. In case treatment without gated radiation therapy for the target with motion of 2 cm, the pass rate was less than 96%. But it was greater than 99% when gated radiation therapy was used. However gated radiation therapy was used for the target with motion greater than 4 cm, the pass rate could not be greater than 97% when gating window was set as 30¡­70%. But when the gating window set as 40¡­60%, the pass rate was greater than 99%.
KEYWORD
Gated radiation therapy, Gamma index, I¡¯m MatriXX, Gating window, Motion range
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed