KMID : 1038720140250020110
|
|
Progress in Medical Physics 2014 Volume.25 No. 2 p.110 ~ p.115
|
|
Comparison of the Dose Distributions with Beam Arrangements in the Stereotactic Body Radiotherapy (SBRT) for Primary Lung Cancer
|
|
Yea Ji-Woon
|
|
Abstract
|
|
|
To compare 2 beam arrangements, circumferential equally angles (EA) beams or partially angles (PA) beams for stereotactic body radiation therapy (SBRT) of primary lung cancer for intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) delivery techniques with respect to target, ipsilateral lung, contralateral lung, and organs-at-risk (OAR) dose-volume metrics, as well as treatment delivery efficiency. Data from 12 patients, four treatment plans were generated per data sets (IMRTEA, IMRTPA, VMATEA, VMATPA). The prescribed dose (PD) was 60 Gy in 4 fractions to 95% of the planning target volume (PTV) for a 6-MV photon beam. When compared with the IMRT and VMAT treatment plan for 2 beams, conformity index, homogeneity index, high dose spillage, D2 cm (Dmax at a distance ¡Ã2 cm beyond the PTV), R50 (ratio of volume circumscribed by the 50% isodose line and the PTV), resulted in similar. But Dmax of the Organ at risk (OAR), spinal cord, trachea, resulted in differ between four treatment plans. Especially HDSlocation showed big difference in 21.63% vs. 26.46%.
|
|
KEYWORD
|
|
Dose distribution, Equal angle, Partial angle, Stereotatic body radiotherapy (SBRT)
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|