KMID : 1038720140250030143
Progress in Medical Physics
2014 Volume.25 No. 3 p.143 ~ p.150
An Experimental Method for the Scatter Correction of MV Images Using Scatter to Primary Ratios (SPRs)
Jeon Ho-Sang

Park Dahl
Kim Dong-Hyun
Lee Ju-Hye
Ki Yong-kan
Lee Ja-Yeong
Nam Ji-ho
Kim Won-Taek
Kim Dong-won
Abstract
In general radiotherapy, mega-voltage (MV) x-ray images are widely used as the unique method to verify radio-therapeutic fields. But, the image quality of MV images is much lower than that of kilo-voltage x-ray images due to scatter interactions. Since 1990s, studies for the scatter correction have performed with digital-based MV imaging systems. In this study, a novel method for the scatter correction is suggested using scatter to primary ratio (SPR), instead of conventional methods such as digital image processing or scatter kernel calculations. We measured two MV images with and without a solid water phantom describing a patient body with given imaging conditions, and calculated un-attenuated ratios. Then, we obtained SPR distributions for the scatter correction. For experimental validation, a line-pair (LP) phantom using several Al bars and a clinical pelvis MV image was used. As the result, scatter signals of the LP phantom image were successfully reduced so that original density distribution of the phantom was restored. Moreover, image contrast values increased after SPR correction at all ROIs of the clinical image. The mean value of increases was 48%. The SPR correction method suggested in this study has high reliability because it is based on actually measured data. Also, this method can be easily adopted in clinics without additional cost. We expected that the SPR correction can be an effective method to improve the quality of MV image guided radiotherapy.
KEYWORD
Scatter to primary ratio, MV image, Scatter correction, Image guided radiotherapy
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed