Progress in Medical Physics
2014 Volume.25 No. 4 p.193 ~ p.198
Qualitative Evaluation of 2D Dosimetry System for Helical Tomotherapy
Ma Sun-Young

Jeung Tae-Sig
Lim Sang-Wook
Shim Jang-Bo
Abstract
The purpose of this study is to see the feasibility of the newly developed 2D dosimetry system using phosphor screen for helical tomotherapy. The cylindrical water phantom was fabricated with phosphor screen to emit the visible light during irradiation. There are three types of virtual target, one is one spot target, another is C-shaped target, and the other is multiple targets. Each target was planned to be treated at 10 Gy by treatment planning system (TPS) of tomotherapy. The cylindrical phantom was placed on the tomotherapy table and irradiated as calculations of the TPS. Every frame which acquired by CCD camera was integrated and the doses were calculated in pixel by pixel. The dose distributions from the fluorescent images were compared with the calculated dose distribution from the TPS. The discrepancies were evaluated as gamma index for each treatment. The curve for dose rate versus pixel value was not saturated until 900 MU/min. The 2D dosimetry using the phosphor screen and the CCD camera is respected to be useful to verify the dose distribution of the tomotherapy if the linearity correction of the phosphor screen improved.
KEYWORD
Tomotherapy, Phosphor screen, Dose distribution, Blur kernel, Gamma index