KMID : 1038720140250040225
Progress in Medical Physics
2014 Volume.25 No. 4 p.225 ~ p.232
Characteristics of Photon Beam through a Handmade Build-Up Modifier as a Substitute of a Bolus
Kim Sung-Joon

Lee Seoung-Jun
Moon Su-Ho
Seol Ki-Ho
Lee Jeong-Eun
Abstract
We evaluated the effect of scatter on a build-up region based on the measured percent depth dose (PDD) of high-energy photon beams that penetrated a handmade build-up modifier (BM) as a substitute of bolus. BM scatter factors (SBM) were calculated based on the PDDs of photon beams that penetrated through the BM. The calculated SBM values were normalized to 1 at the square field side (SFS) of 30 mm without a BM. For the largest SFS (200 mm), the SBM values for a 6-MV beam were 1.331, 1.519, 1.598, 1.641, and 1.657 for the corresponding BM thickness values. For a 10-MV beam, the SBM values were 1.384, 1.662, 1.825, 1.913, and 2.001 for the corresponding BM thickness values. The BM yielded 76% of the bolus efficiency. We expect BM to become useful devices for deep-set patient body parts to which it is difficult to apply a bolus.
KEYWORD
Build-up modifier (BM), BM scatter factor (SBM), Percent depth dose (PDD), Surface dose