Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1038720140250040248
Progress in Medical Physics
2014 Volume.25 No. 4 p.248 ~ p.254
Study on Generator Control for a Small X-Ray Tube
Lee Soon-hyouk

Ji Yun-Seo
Choi Sang-Gyu
Lee Re-Na
Abstract
The purpose of this study is to develope a control system for a small X-ray tube generator and investigate control methods for the X-ray generator. The small X-ray tube was developed for electronic brachytherapy, and thus, the new control method should be investigated, if the small X-ray tube is used for the imaging system. The Axxent S700 X-ray tube and the XF060NZZ485 high voltage generator were used to compose a X-ray imaging system and control board was developed by using AT90CAN128 MCU. The two control methods were investigated after tube voltage was set to 50 kV, one was filament current control method and the other was beam current control method. The former was subdivided into two methods according to the filament heating time, the 5 and the 10 seconds respectively. In the filament current method, the beam current did not rise up to the desired value, if the filament current had not been maintained for at least 10 seconds. The onset filament currents to generate beam current were varied from 1,300 to 1,350 mA and over 5 seconds were needed in order to reach the desired tube current value after beam current was generated. However, in the tube current control method, the beam current reached to the desired value without any time delay with the filament current of 1,500 mA. In this study, we found that the beam current control method was appropriate for the use of small X-ray tube developed for brachytherapy in the X-ray imaging system.
KEYWORD
Intra-oral X-ray tube, X-ray generator control, Beam current control
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed