KMID : 1038720140250040281
Progress in Medical Physics
2014 Volume.25 No. 4 p.281 ~ p.287
Isocenter Check and QA of Tactic Radiosurgery Devices Using EPID
Shin Kyo-Chul

Choi Sang-Gyu
Kim Jung-Kee
Jeong Dong-Hyeok
Abstract
The Winston Lutz test, which checks the accuracy of the isocenter for stereotactic radiosurgery (SRS), was performed with the commercial electronic portal imaging device (EPID). The usual Winston Lutz test with film was also performed for comparison with the test with EPID. The maximum difference in isocenter between the two methods was 0.32 mm. The Winston Lutz test using EPID is practical as it can reduce time and avoid human errors compared to the test with film.
KEYWORD
SRS, EPID, Winston Lutz Test, Quality assurance