KMID : 1038720160270010008
|
|
Progress in Medical Physics 2016 Volume.27 No. 1 p.8 ~ p.13
|
|
Analysis of Beam Hardening of Modulation Layers for Dual Energy Cone-beam CT
|
|
Ahn So-Hyun
Cho Sam-Ju Keum Ki-Chang Choi Sang-Gyu Lee Re-Na
|
|
Abstract
|
|
|
Dual energy cone-beam CT can distinguish two materials with different atomic compositions. The principle of dual energy cone-beam CT based on modulation layer is that higher energy spectrum can be acquired at blocked x-ray window. To evaluate the possibility of modulation layer based dual energy cone-beam CT, we analyzed x-ray spectrum for various thicknesses of modulation layers by Monte Carlo simulation. To compare with the results of simulation, the experiment was performed on prototype cone-beam CT for 50~100 kVp with CdTe XR-100T detector. As the result of comparing, the mean energy of energy spectrum for 80 kVp are well matched with that of simulation. The mean energy of energy spectrum for 80 and 120 kVp were increased as 1.67 and 1.52 times by 2.0 mm modulation layer, respectively. We realized that the virtual dual energy x-ray source can be generated by modulation layer.
|
|
KEYWORD
|
|
Dual energy cone-beam CT, Beam hardening effect, Modulation layer, Monte Carlo simulation
|
|