KMID : 1038720160270030169
Progress in Medical Physics
2016 Volume.27 No. 3 p.169 ~ p.174
Treatment Plan Delivery Accuracy of the ViewRay System in Two-Headed Mode
Park Jong-Min

Park So-Yeon
Wu Hong-Gyun
Kim Jung-In
Abstract
The aim of this study is to investigate the delivery accuracy of intensity-modulated radiation therapy (IMRT) plans in the two-headed mode of the ViewRay¢â system in comparison with that of the normal operation treatment plan of the machine. For this study, a total of eight IMRT plans and corresponding verification plans were generated (four head and neck, two liver, and two prostate IMRT plans). The delivered dose distributions were measured using ArcCHECK¢â with the insertion of an ionization chamber. We measured the delivered dose distributions in three-headed mode (normal operation of the machine), two-headed mode with head 1 disabled, two-headed mode with head 2 disabled, and two-headed mode with head 3 disabled. Therefore, a total of four measurements were performed for each IMRT plan. The global gamma passing rates (3%/3 mm) in three-headed mode, head 1 disabled, head 2 disabled, and head 3 disabled were 99.9¡¾0.1%, 99.8¡¾0.3%, 99.6¡¾0.7%, and 99.7¡¾0.4%, respectively. The difference in the gamma passing rates of the three- and two-headed modes was insignificant. With 2%/2 mm, the rates were 96.6¡¾3.6%, 97.2¡¾3.5%, 95.7¡¾6.2%, and 95.5¡¾4.3%, respectively. Between three-headed mode and head 3 disabled, a statistically significant difference was observed with a p-value of 0.02; however, the difference was minimal (1.1%). The chamber readings showed differences of approximately 1% between three- and two-headed modes, which were minimal. Therefore, the treatment plan delivery in the two-headed mode of the ViewRay¢â system seems accurate and robust.
KEYWORD
MRI-guided radiation therapy system, Two-headed mode, Co-60, Intensity modulated radiation therapy
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed