Development Treatment Planning System Based on Monte-Carlo Simulation for Boron Neutron Capture Therapy
Kim Moo-Sub

Kubo Kazuki
Monzen Hajime
Yoon Do-Kun
Shin Han-Back
Kim Sun-Mi
Suh Tae-Suk
Abstract
The purpose of this study is to develop the treatment planning system (TPS) based on Monte-Carlo simulation for BNCT. In this paper, we will propose a method for dose estimation by Monte-Carlo simulation using the CT image, and will evaluate the accuracy of dose estimation of this TPS. The complicated geometry like a human body allows defining using the lattice function in MCNPX. The results of simulation such as flux or energy deposition averaged over a cell, can be obtained using the features of the tally provided by MCNPX. To assess the dose distribution and therapeutic effect, dose distribution was displayed on the CT image, and dose volume histogram (DVH) was employed in our developed system. The therapeutic effect can be efficiently evaluated by these evaluation tool. Our developed TPS could be effectively performed creating the voxel model from CT image, the estimation of each dose component, and evaluation of the BNCT plan.
KEYWORD
BNCT, Monte Carlo, MCNPX, Treatment planning system