KMID : 1038720170280010022
Progress in Medical Physics
2017 Volume.28 No. 1 p.22 ~ p.26
Development of an Analytic Software Using Pencil Beam Scanning Proton Beam
Jeong Seong-Hoon

Yoon Myong-Geun
Chung Kwang-Zoo
Han Young-Yih
Lim Do-Hoon
Choi Doo-Ho
Abstract
We have developed an analytic software that can easily analyze the spot position and width of proton beam therapy nozzles in a periodic quality assurance. The developed software consists of an image processing method that conducts an analysis using center-of-spot geometry and a Gaussian fitting method that conducts an analysis through Gaussian fitting. By using the software, an analysis of 210 proton spots with energies 150, 190, and 230 MeV showed a deviation of approximately 3% from the mean. The software we developed to analyze proton spot positions and widths provides an accurate analysis and reduces the time for analysis.
KEYWORD
Proton therapy, Quality assurance, Proton spot analysis, Spot position and width