Development of an Analytic Software Using Pencil Beam Scanning Proton Beam
|
|
Jeong Seong-Hoon
Yoon Myong-Geun Chung Kwang-Zoo Han Young-Yih Lim Do-Hoon Choi Doo-Ho
|
|
Abstract
|
|
|
We have developed an analytic software that can easily analyze the spot position and width of proton beam therapy nozzles in a periodic quality assurance. The developed software consists of an image processing method that conducts an analysis using center-of-spot geometry and a Gaussian fitting method that conducts an analysis through Gaussian fitting. By using the software, an analysis of 210 proton spots with energies 150, 190, and 230 MeV showed a deviation of approximately 3% from the mean. The software we developed to analyze proton spot positions and widths provides an accurate analysis and reduces the time for analysis.
|
|
KEYWORD
|
|
Proton therapy, Quality assurance, Proton spot analysis, Spot position and width
|
|