KMID : 1038720180290040164
Progress in Medical Physics
2018 Volume.29 No. 4 p.164 ~ p.172
Analysis of Small-Field Dosimetry with Various Detectors
Park So-Yeon

Choi Byeong-Geol
Lee Dong-Myung
Jang Na-Young
Abstract
We evaluated the performance of various detectors for small-field dosimetry with field sizes defined by a high-definition (HD) multileaf collimator (MLC) system. For small-field dosimetry, diodes referred to as ¡°RAZOR detectors,¡± MOSFET detectors, and Gafchromic EBT3 films were used in this study. For field sizes less than 1¡¿1 cm2, percent depth doses (PDDs) and lateral profiles were measured by diodes, MOSFET detectors, and films, and absolute dosimetry measurements were conducted with MOSFET detectors. For comparison purposes, the same measurements were carried out with a field size of 10¡¿10 cm2. The dose distributions were calculated by the treatment planning system Eclipse. A comparison of the measurements with calculations yielded the percentage differences. With field sizes less than 1¡¿1 cm2, it was shown that most of the percentage difference values were within 5% for 6-MV and 15-MV photon beams with the use of diodes. The measured lateral profiles were well matched with those calculated by Eclipse as the field sizes increased. Except for the depths of 0.5 cm and 20 cm, there was agreement in terms of the absolute dosimetry within 10% when MOSFET detectors were used. There was good agreement between the calculations and measurements conducted using diodes and EBT films. Both diode detectors and EBT3 films were found to be appropriate options for relative measurements of PDDs and for lateral profiles.
KEYWORD
Analysis of Small-Field Dosimetry with Various Detectors
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed