Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1038720190300040120
Progress in Medical Physics
2019 Volume.30 No. 4 p.120 ~ p.127
Feasibility Study of Mobius3D for Patient-Specific Quality Assurance in the Volumetric Modulated Arc Therapy
Lee Chang-Yeol

Kim Woo-Chul
Kim Hun-Jeong
Lee Jeong-Shim
Huh Hyun-Do
Abstract
Purpose: This study was designed to evaluate the dosimetric performance of Mobius3D by comparison with an aSi-based electronic portal imaging device (EPID) and Octavius 4D, which are conventionally used for patient-specific prescription dose verification.

Methods: The study was conducted using nine patients who were treated by volumetric modulated arc therapy. To evaluate the feasibility of Mobius3D for prescription dose verification, we compared the QA results of Mobius3D to an aSi-based EPID and the Octavius 4D dose verification methods. The first was the comparison of the Mobius3D verification phantom dose, and the second was to gamma index analysis.

Results: The percentage differences between the calculated point dose and measurements from a PTW31010 ion chamber were 1.6%¡¾1.3%, 2.0%¡¾0.8%, and 1.2%¡¾1.2%, using collapsed cone convolution, an analytical anisotropic algorithm, and the AcurosXB algorithm respectively. The average difference was found to be 1.6%¡¾0.3%. Additionally, in the case of using the PTW31014 ion chamber, the corresponding results were 2.0%¡¾1.4%, 2.4%¡¾2.1%, and 1.6%¡¾2.5%, showing an average agreement within 2.0%¡¾0.3%. Considering all the criteria, the Mobius3D result showed that the percentage dose difference from the EPID was within 0.46%¡¾0.34% on average, and the percentage dose difference from Octavius 4D was within 3.14%¡¾2.85% on average.

Conclusions: We conclude that Mobius3D can be used interchangeably with phantom-based dosimetry systems, which are commonly used as patient-specific prescription dose verification tools, especially under the conditions of 3%/3 mm and 95% pass rate.
KEYWORD
Mobius3D, Patient-specific quality assurance, Volumetric modulated arc therapy, Gamma index analysis
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed