KMID : 1038720210320040116
Progress in Medical Physics
2021 Volume.32 No. 4 p.116 ~ p.121
Evaluating the Effects of Dose Rate on Dynamic Intensity-Modulated Radiation Therapy Quality Assurance
Kim Kwon-Hee

Back Tae-Seong
Chung Eun-Ji
Suh Tae-Suk
Sung Won-Mo
Abstract
Purpose: To investigate the effects of dose rate on intensity-modulated radiation therapy (IMRT) quality assurance (QA).

Methods: We performed gamma tests using portal dose image prediction and log files of a multileaf collimator. Thirty treatment plans were randomly selected for the IMRT QA plan, and three verification plans for each treatment plan were generated with different dose rates (200, 400, and 600 monitor units [MU]/min). These verification plans were delivered to an electronic portal imager attached to a Varian medical linear accelerator, which recorded and compared with the planned dose. Root-mean-square (RMS) error values of the log files were also compared.

Results: With an increase in dose rate, the 2%/2-mm gamma passing rate decreased from 90.9% to 85.5%, indicating that a higher dose rate was associated with lower radiation delivery accuracy. Accordingly, the average RMS error value increased from 0.0170 to 0.0381 cm as dose rate increased. In contrast, the radiation delivery time reduced from 3.83 to 1.49 minutes as the dose rate increased from 200 to 600 MU/min.

Conclusions: Our results indicated that radiation delivery accuracy was lower at higher dose rates; however, the accuracy was still clinically acceptable at dose rates of up to 600 MU/min.
KEYWORD
Intensity modulated radiation therapy, Quality assurance, Dose rate, Multileaf collimator