Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1038720210320040145
Progress in Medical Physics
2021 Volume.32 No. 4 p.145 ~ p.152
Measurements of Neutron Activation and Dose Rate Induced by High-Energy Medical Linear Accelerator
Kwon Na-Hye

Jang Young-Jae
Kim Jin-Sung
Kim Kum-Bae
Yoo Jae-Ryong
Ahn So-Hyun
Kim Dong-Wook
Choi Sang-Hyoun
Abstract
Purpose: During the treatments of cancer patients with a linear accelerator (LINAC) using photon beams with energies ¡Ã8 MV, the components inside the LINAC head get activated through the interaction of photonuclear reaction (¥ã, n) and neutron capture (n, ¥ã). We used spectroscopy and measured the dose rate for the LINAC in operation after the treatment ended.

Methods: We performed spectroscopy and dose rate measurements for three units of LINACs with a portable high-purity Germanium (HPGe) detector and a survey meter. The spectra were obtained after the beams were turned off. Spectroscopy was conducted for 3,600 seconds, and the dose rate was measured three times. We identified the radionuclides for each LINAC.

Results: According to gamma spectroscopy results, most of the nuclides were short-lived radionuclides with half-lives of 100 days, except for 60Co, 65Zn, and 181W nuclides. The dose rate for three LINACs obtained immediately in front of the crosshair was in the range of 0.113 to 0.129 ¥ìSv/h. The maximum and minimum dose rates measured on weekends were 0.097 ¥ìSv/h and 0.092 ¥ìSv/h, respectively. Compared with the differences in weekday data, there was no significant difference between the data measured on Saturday and Sunday.

Conclusions: Most of the detected radionuclides had half-lives <100 days, and the dose rate decreased rapidly. For equipment that primarily used energies ¡Â10 MV, when the equipment was transferred after at least 10 minutes after shutting it down, it is expected that there will be little effect on the workers¡¯ exposure.
KEYWORD
Spectroscopy, Activation, Dose rate, Linear accelerator, Radionuclide
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed