KMID : 1038720210320040159
Progress in Medical Physics
2021 Volume.32 No. 4 p.159 ~ p.164
Assessing Commercial CLEANBOLUS Based on Silicone for Clinical Use
Son Jae-Man

Jung Seong-Moon
Park Jong-Min
Choi Chang-Heon
Kim Jung-In
Abstract
Purpose: We investigated the properties of CLEANBOLUS based on silicone with suitable characteristics for clinical use.

Methods: We evaluated the characteristics of CLEANBOLUS and compared the results with the commercial product (Super-Flex bolus). Also, we conducted physical evaluations, including shore hardness, element composition, and elongation break. Transparency was investigated through the measured absorbance within the visible region (400?700 nm). Also, dosimetric characteristics were investigated with surface dose and beam quality. Finally, the volume of unwanted air gap was investigated based on computed tomography images for breast, chin, and nose using Super-Flex bolus and CELANBOLUS.

Results: CLEANBOLUS showed excellent physical properties for a low shore hardness (000?35) and elongation break (>1,000%). Additionally, it was shown that CLEANBOLUS is more transparent than Super-Flex bolus. Dosimetric results obtained through measurement and calculation have an electron density similar to water in CLEANBOLUS. Finally, CLEANBOLUS showed that the volume of unwanted air gap between the phantom and each bolus is smaller than Super-Flex bolus for breast, chin, and nose.

Conclusions: The physical properties of CLEANBOLUS, including excellent adhesive strength and lower shore hardness, reduce unwanted air gaps and ensure accurate dose distribution. Therefore, it would be an alternative to other boluses, thus improving clinical use efficiency.
KEYWORD
Bolus, Transparency, Surface dose, Air gap