Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1038720220330040072
Progress in Medical Physics
2022 Volume.33 No. 4 p.72 ~ p.79
Proficiency Test for the Dosimetry Audit Service Provider
Yi Chul-Young

Kim In-Jung
Park Jong-In
Kim Yun-Ho
Seong Young-Min
Abstract
Purpose: The proficiency test was conducted to assess the performance of the dosimetry audit service provider in the readout practice of the dose delivered to patients in medical institutions.

Methods: A certain amount of the absorbed dose to water for the high-energy X-ray from the medical linear accelerator (LINAC) installed in the Korea Research Institute of Standards and Science (KRISS) was delivered to the postal dose audit package given by the dosimetry audit service provider, in which the radio-photoluminescence (RPL) glass dosimeters were mounted. The dosimetry audit service provider read the RPL glass dosimeters and sent the readout dose value with its uncertainty to KRISS. The performance of the dosimetry audit service provider was evaluated based on the En number given in ISO/IEC 17043:2010.

Results: The evaluated En number was ?0.954. Based on the ISO/IEC 17043, the performance of the dosimetry service provider is ¡°satisfactory.¡±

Conclusions: As part of the conformity assessment, the KRISS performed the proficiency test over the postal dose audit practice run by the dosimetry audit service provider. The proficiency test is in line with confirming the traceability of the medical institutions to the primary standard of absorbed dose to the water of the KRISS and ensuring the confidence of the dosimetry audit service provider.
KEYWORD
Proficiency test, Dosimetry audit, Absorbed dose to water, Primary standard, Traceability
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed