Progress in Medical Physics
2022 Volume.33 No. 4 p.108 ~ p.113
Feasibility Test of Flat-Type Faraday Cup for Ultrahigh-Dose-Rate Transmission Proton Beam Therapy
Pak Sang-Il

Cho Sung-Koo
An Seo-Hyeon
Jeong Seong-Hoon
Shin Dong-Ho
Lim Young-Kyung
Jeong Jong-Hwi
Kim Hak-Soo
Lee Se-Byeong
Abstract
Purpose: Proton therapy has been used for optimal cancer treatment by adapting its Bragg-peak characteristics. Recently, a tissue-sparing effect was introduced in ultrahigh-dose-rate (FLASH) radiation; the high-energy transmission proton beam is considered in proton FLASH therapy. In measuring high-energy/ultrahigh-dose-rate proton beam, Faraday Cup is considered as a dose-rate- independent measurement device, which has been widely studied. In this paper, the feasibility of the simply designed Faraday Cup (Poor Man¡¯s Faraday Cup, PMFC) for transmission proton FLASH therapy is investigated.

Methods: In general, Faraday cups were used in the measurement of charged particles. The simply designed Faraday Cup and Advanced Markus ion chamber were used for high-energy proton beam measurement in this study.

Results: The PMFC shows an acceptable performance, including accuracy in general dosimetric tests. The PMFC has a linear response to the dose and dose rate. The proton fluence was decreased with the increase of depth until the depth was near the proton beam range. Regarding secondary particles backscatter from PMFC, the effect was negligible.

Conclusions: In this study, we performed an experiment to investigate the feasibility of PMFC for measuring high-energy proton beams. The PMFC can be used as a beam stopper and secondary monitoring system for transmission proton beam FLASH therapy.
KEYWORD
Proton beam therapy, Faraday cup, FLASH