Feasibility of a Linear Diode Array Detector for Commissioning of a Radiotherapy Planning System
Hong Seung-Mo

Lee Ui-Seob
Kim Sung-Woo
Goh Young-Moon
Abstract
Purpose: Although ionization chambers are widely used to measure beam commissioning data, point-by-point measurements of all the profiles with various field size and depths are time- consuming tasks. As an alternative, we investigated the feasibility of a linear diode array for commissioning a treatment planning system.

Methods: The beam data of a Varian TrueBeam¢ç radiotherapy system at 6 and 10 MV with/without a flattening filter were measured for commissioning of an Eclipse Analytical Anisotropic Algorithm (AAA) ver.15.6. All of the necessary beam data were measured using an IBA CC13 ionization chamber and validated against Varian ¡°Golden Beam¡± data. After validation, the measured CC13 profiles were used for commissioning the Eclipse AAA (AAACC13). In addition, an IBA LDA-99SC linear diode array detector was used to measure all of the beam profiles and for commissioning a separate model (AAALDA99). Finally, the AAACC13 and AAALDA99 dose calculations for each of the 10 clinical plans were compared.

Results: The agreement of the CC13 profiles with the Varian Golden Beam data was confirmed within 1% except in the penumbral region, where ¡Â2% of a discrepancy related to machine- specific jaw calibration was observed. Since the volume was larger for the CC13 chamber than for the LDA-99SC chamber, the penumbra widths were larger in the CC13 profiles, resulting in ¡Â5% differences. However, after beam modeling, the penumbral widths agreed within 0.1 mm. Finally the AAA LDA99 and AAACC13 dose distributions agreed within 1% for all voxels inside the body for the 10 clinical plans.

Conclusions: In conclusion, the LDA-99SC diode array detector was found to be accurate and efficient for measuring photon beam profiles to commission treatment planning systems.
KEYWORD
Radiotherapy, Photon beam commissioning, Treatment planning system, Ionization chamber, Diode array detector