2013 Volume.3 No. 4 p.198 ~ p.212
Analytical Performance of Wako and Sekisui Clinical Chemistry Assays on Hitachi LABOSPECT 008
Chae Hyo-Jin

Park Hae-Il
Kim Myung-Shin
Lee Je-Hoon
Kim Yong-Goo
Abstract
Background: We evaluated the analytical performance of Wako assays for albumin, alanine aminotransferase (ALT), aspartate aminotransferase (AST), blood urea nitrogen (BUN), complement C3 and C4, calcium, creatine kinase (CK), C-reactive protein (CRP), direct bilirubin (DBIL), iron, gamma-glutamyl transferase (GGT), HDL cholesterol (HDLC), inorganic phosphorus (IP), LDL cholesterol (LDLC), total bilirubin (TBIL), total protein (TP), and uric acid (UA), as well as the performance of Sekisui assays for albumin, BUN, calcium, CRP, HDLC, IP, LDLC, TP, and UA by using Hitachi LABOSPECT 008 (Hitachi High-Tech Co., Japan).

Methods: Liquid Assayed Multiqual Control (Bio-Rad Laboratories, USA) and pooled patients' sera were analyzed for 20 days. Wako linearity material (Wako Pure Chemical Industries, Ltd., Japan) and Sysmex Interference Check A Plus kit (Sysmex Co., Japan) were used to test linearity and interference, respectively. Concentrations of the target analytes were measured using Hitachi LABOSPECT 008 in 100 residual patient specimens and compared to those in Pureauto S series reagent (Sekisui Medical, Japan), which were measured using Hitachi 7600 (Hitachi High-Tech Co., Japan).

Results: Total coefficients of variation (CVs) for the tested analytes were 0.91-9.26% in Wako and 1.04-7.46% in Sekisui assays. Linearity was demonstrated up to the highest concentration within the analytical range in all the assays except for Wako albumin and Sekisui TP. Wako and Sekisui albumin, BUN, CRP, HDLC, and LDLC assays, and in Wako C3, C4, calcium, and UA assays showed no interference with the test concentrations used. All the tested assays, except for Wako AST, LDLC, and TP, and Sekisui calcium and TP, demonstrated comparability with comparative method for at least one medical decision level.

Conclusions: Our study results showed that the analytical performances of Wako and Sekisui chemistry assays evaluated using Hitachi LABOSPECT 008 had appropriate analytical performance for clinical use.
KEYWORD
Chemistry Techniques, Analytical, Evaluation Studies, Clinical Chemistry Tests, Hitachi LABOSPECT 008