Evaluation of Centaur Syphilis, Immulite Syphilis, and Mediace TPLA for Detecting Treponemal Antibodies
Seo Dong-Hee

Whang Dong-Hee
Joo Shin-Young
Choi Hyen-Hee
Abstract
Background: We evaluated the efficacy of two chemiluminescence immunoassays that detect treponemal antibodies, Centaur Syphilis and Immulite Syphilis, in comparison with Mediace Treponema pallidum latex agglutination (TPLA).

Methods: The study was conducted in two phases. In the first phase, we tested 1,147 serum samples that were sequentially submitted for routine syphilis serology. In the second phase, we tested a panel of 119 frozen serum samples that had previously tested positive by Mediace RPR. The kappa value, total agreement percentage, and sensitivity and specificity were analyzed.

Results: Of the 1,147 random samples, 24 (2.09%) tested positive with Centaur Syphilis, 16 (1.39%) with Immulite Syphilis, and 19 (1.66%) with Mediace TPLA. Of the 119 Mediace RPR-positive samples, 103 (86.6%) tested positive with Centaur Syphilis, 101 (84.9%%) with Immulite Syphilis, and 105 (88.2%) with Mediace TPLA. The percent agreements (kappa values) were 98.8% (0.934) between Centaur Syphilis and Mediace TPLA, 99.0% (0.94) between Immulite Syphilis and Mediace TPLA, and 99.2% (0.955) between Centaur Syphilis and Immulite Syphilis. To measure the sensitivity and specificity of each treponemal test, samples showing agreement in three or four of the tests (three treponemal tests and Mediace-RPR) were regarded as true positive (n=117) or true negative (n=1,142). The respective values for sensitivity and specificity were 100% and 99.6% for Centaur Syphilis, 98.3% and 100% for Immulite Syphilis, and 99.2% and 99.7% for Mediace TPLA.

Conclusions: Results from the three treponemal assays were in good agreement. Greater sensitivity of Centaur Syphilis and greater specificity of Immulite Syphilis were suggested.
KEYWORD
Syphilis, Treponema pallidum, Chemiluminescence immunoassay, Treponemal test, Non-treponemal test