KMID : 1100820200100010046
Laboratory Medicine Online
2020 Volume.10 No. 1 p.46 ~ p.51
Evaluation of Analytical Performance of an Automated Glycated Hemoglobin Analyzer, HLC-723 G11
Chung Yoo-Na

Yun Seung-Gyu
Cho Yun-Jung
Abstract
Background: Glycated hemoglobin (HbA1c) is a key biomarker for the monitoring of glycemic balance in patients with diabetes. The aim of this study was to evaluate the performance of a new system, the Tosoh HLC-723 G11 analyzer (Tosoh Corporation, Japan), compared to that of two routine diagnostic testing systems, Tosoh G8 (Tosoh Corporation) and Capillarys 2 Flex Piercing (Sebia, France).

Methods: Tosoh G11 was evaluated for precision, linearity, and carry-over, according to the Clinical and Laboratory Standard Institute's guidelines. Test results from clinical samples were compared between Tosoh G11 and the routine testing systems, Tosoh G8 and Capillarys 2 Flex Piercing.

Results: With respect to the precision of Tosoh G11, the test results for low- and high-concentration controls showed a coefficient of variation of less than 1.1%. Furthermore, the new device exhibited good linearity for HbA1c values ranging from 3.4% to 18.8%, and carry-over was not observed. HbA1c results for Tosoh G11 (N=143) correlated well with those for Tosoh G8 (r=0.9971) and Capillarys 2 Flex Piercing (r=0.9918).

Conclusions: Tosoh G11 demonstrated reliable analytical performance with good precision and linearity, and no carry-over results. In addition, its results were comparable to those of the existing instruments. Thus, the results of this evaluation suggest that Tosoh G11 is suitable for the routine diagnostic testing of HbA1c levels in clinical chemistry laboratories.
KEYWORD
Hemoglobin A1c, Tosoh G11, Analytical performance