KMID : 1100820200100030197
Laboratory Medicine Online
2020 Volume.10 No. 3 p.197 ~ p.201
Performance Evaluation of Serum IgD Quantification by the SPAPLUS Turbidimetric Analyzer and Determination of a Reference Interval of IgD in the Korean Population
Cho Eun-Hye

Park Hyun-Kyung
Park Hyung-Doo
Abstract
Background: IgD is an immunoglobulin isotype that accounts for <1% of the total serum immunoglobulin, and IgD level changes can be observed in specific clinical situations. Herein, we evaluated the analytical performances of serum IgD quantification by the SPAPLUS turbidimetric analyzer (The Binding Site, UK), and determined a reference interval of IgD in the Korean population.

Methods: Precision, linearity and comparison with BNII (Siemens Healthineers, Germany) were assessed, and a reference interval was established according to the Clinical and Laboratory Standards Institute guidelines.

Results: Repeatability ranged from 0.78% to 2.14%, and the within-laboratory precision ranged from 4.10% to 5.40% for three concentrations. The coefficient of determination (R2) was 0.9996 for the analytical measurement range. The IgD concentration showed a correlation between SPAPLUS and BNII (y=0.890x-1.852, r=0.984). However, IgD concentrations tended to be lower in SPAPLUS compared to those in BNII at concentrations >100 mg/L. The established upper limit of the reference interval was 53.6 mg/L.

Conclusions: The SPAPLUS shows good analytical performance for serum IgD quantification.
KEYWORD
Analytical performance, IgD, Monoclonal gammopathy, Reference interval, Turbidimetry