Korean Journal of Nuclear Medicine Technology
2019 Volume.23 No. 2 p.43 ~ p.50
Comparison Analysis of The results of IRMA Test among Different Equipment According to Algorithm change
Kim Jung-In

Kwon Won-Hyun
Lee Kyung-Jae
Abstract
Purpose:The principle of nuclear medicine test is divided into two main categories: competition(radioimmunoassay, RIA) and noncompetitive reaction(Immunoradiometric assay, IRMA). It is known that the curve fitting method, which is commonly used in inspection field, uses Spline interpolation in RIA method and Linear interpolation method in IRMA method. Among them, the insulin test using the IRMA test showed a significant difference, especially at low concentrations, despite the same algorithm of linear interpolation between fully automated radio immunoassay analyzers. In this study, we aim to obtain results from applying two different of algorithm using fully automated radio immunoassay analyzers including Gamma pro, Gamma 10, Cobra, and SR300.

Materials and Methods: A total of 30 test samples were selected for the test of TSH, ferritin, C-peptide, and insulin serum levels. Test was performed by IRMA method. We compared the difference in the results of applying the linear interpolation method and the spline interpolation method to Gamma Pro, Gamma 10, Cobra, and SR300 equipment.

Results: Two-way ANOVA was used for statistical analysis. The significance level was applied as P <0.05. The results of TSH, ferritin, C-peptide, and insulin tests were compared between the fully automated radio immunoassay analyzers. There was a significant difference between ferritin, C-peptide, and insulin serum levels(P<0.001). TSH didn't show any significant different between the devices(P=0.29). In the difference between linear and spline interpolation, there was no significant difference between insulin test(P=0.08), TSH test(P=0.81), and Ferritin test(P=0.06). However, C-peptide test showed a significant difference(P=0.03). Especially, the insulin test showed significant difference in lower ranges. As a result of comparing and analyzing the difference between the two interpolation methods, the devices in the low concentration group showed significant difference(P<0.001).

Conclusion: In case of new equipment in the laboratory it is necessary to recognize that there is a difference in the curve fitting method for each automated radio immunoassay analyzers in the low concentration area when the principle of inspection is IRMA method.
KEYWORD
IRMA method, Linear interpolation, Spline interpolation