Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1147920210040010010
Medical Biological Science and Engineering
2021 Volume.4 No. 1 p.10 ~ p.19
Effect of intravenous fluid warming devices on intraoperative core temperature during maintenance fluid therapy: a prospective randomized controlled trial
Park Jeong-Wook

Jung Ki-Tae
Jang Bo-Hyun
Kim Sang-Hun
Abstract
Background: We investigated the effects of intravenous fluid warmers on intraoperative core temperatures at low flow rates during the maintenance fluid therapy.

Methods: We conducted a prospective, non-blinded, randomized controlled trial in 99 patients, scheduled for elective spinal fusion surgery with more than 3 hours of general anesthesia. They were randomly distributed into groups using Mega Acer Kit (group M; n=30), Ranger (group R; n=32), or ThermoSens (group T; n=32). The infused flow rate of the fluid was determined as one-third of the preoperative fluid deficit (4-2-1 rules) during fasting times, and the third space loss during surgery (2 mL/kg), per hour. The primary outcome was intraoperative final and lowest esophageal temperature (Teso_ Final, lowest Teso).

Results: Teso_Final, lowest Teso, and intraoperative Teso were not significantly different between groups (p=0.512, and p=0.393, p=0.066, respectively). However, the temperature change from baseline Teso to Teso_Final was significantly different between the groups (p=0.044), which of group M were significantly lower than group T (p=0.033). After adjustment with baseline Teso, the differences of least squares means showed the significant differences between groups M and T at 2.5 hours (p=0.020) and 3 hours (p=0.006).

Conclusions: The Mega Acer Kit, Ranger, and ThermoSens have a similar effect on intraoperative core temperatures with the low flow rates, but The Mega Acer Kit is more effective at controlling core temperature than the ThermoSens if the fluid infusion time is over 2.5 hours.
KEYWORD
Body temperature changes, Heating/instrumentation, Hypothermia, Intravenous infusions, Maintenance flow rate
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)