Neurointervention 2016 Volume.11 No. 1 p.24 ~ p.29
|
|
Easy Advancement of a Large-Profile Microcatheter (Excelsior XT27¢â) by Parallel Use of Two Microguidewires For Stent Delivery
|
|
Adrianto Yudhi
Yang Ku-Hyun Koo Hae-Won Park Won-Hyoung Park Jung-Cheol Lee Deok-Hee
|
|
Abstract
|
|
|
Purpose: In the endovascular treatment of cerebral aneurysms, navigating a large-bore microcatheter for delivery of an open-cell stent can be challenging, especially in wide-necked bifurcation aneurysms. We were able to overcome this difficulty by parallel use of two microguidewires through the stent-delivery microcatheter.
Materials and Methods: From December 2014 to April 2015, we treated 15 patients with wide-necked bifurcation aneurysms. For stent delivery, we used a 300-cm 0.014-in microguidewire (Transend), which was placed into the target branch using an exchange technique. A 0.027-in microcatheter (Excelsior XT-27), which was designed for the stent, was advanced over the exchange microguidewire. If we had trouble in advancing the microcatheter over the exchange microguidewire, we inserted a regular microguidewire (Traxcess), into the microcatheter lumen in a parallel fashion. We also analyzed the mechanism underlying microcatheter positioning failure and the success rate of the 'parallel-wire technique'.
Results: Among the 15 cases, we faced with navigation difficulty in five patients. In those five cases, we could advance the microcatheter successfully by applying the parallel-wire technique. There were no procedure-related complications.
Conclusion: Simply by using another microguidewire together with pre-existing microguidewire in a parallel fashion, the stent-delivery microcatheter can be easily navigated into the target location in case of any advancement difficulty.
|
|
KEYWORD
|
|
Cerebral aneurysm, Stent-assisted coiling, Intracranial stenting, Microcatheter, Microguidewire
|
|