Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1155820240190010006
Neurointervention
2024 Volume.19 No. 1 p.6 ~ p.13
A ¡°Radial Ready¡± Tricoaxial Setup for Anterior Circulation Mechanical Thrombectomy: Technical Aspects and Preliminary Results
Stefano Molinaro

Riccardo Russo
Francesco Mistretta
Gaetano Risi
Umberto Amedeo Gava
Mauro Bergui
Abstract
Purpose: Mechanical thrombectomy (MT) is the standard of care for acute ischemic stroke (AIS) due to large vessel occlusion (LVO). The choice of a transradial approach (TRA) for anterior circulation LVOs is still debatable; the use of a specific tricoaxial system could help mitigate numerous issues related to transradial MT.

Materials and Methods: From November 2022 to November 2023, 22 patients underwent TRA-MT for anterior circulation LVOs, both as first-line and rescue from transfemoral approach (TFA) failure, with the same triaxial setup consisting of a 7F introducer sheath, 7F guide catheter, and aspiration catheters ranging from 5.5F to 5F in relation to the occlusion site. Choice of thrombectomy technique was at operator discretion. Patients¡¯ demographic data, clinical presentation, treatment details, complications, rate of crossover to TFA, successful revascularization (modified thrombolysis in cerebral infarction [mTICI] score ¡Ã2b), and good clinical outcome at 3 months (modified Rankin scale [mRS] 0-2) were reported.

Results: Of 20 patients selected, 10 (50%) had occlusion of M1 segment of middle cerebral artery (MCA), 6 (30%) of internal carotid artery (ICA) terminus, and 4 (20%) with M2 MCA occlusions; 12/20 (60%) were right-sided occlusions and 8/20 (40%) were left-sided. The mean National Institutes of Health Stroke Scale score was 9.25 at admission. Successful revascularization to mTICI 2b-3 was achieved in 18/20 patients (90%). Intracranial complications were reported in 2 (10%) patients. Rate of radial artery occlusion at 24 hours was 10,6%; no access-site haemorrhagic complications were reported. Symptomatic intracranial hemorrhage occurred in 2 (10%) patients. mRS score 0-2 at 3 months was 50%.

Conclusion: The high technical effectiveness and good safety profile of this specific tricoaxial setup for TRA-MT in AIS, even for large proximal LVOs, could constitute a viable alternative to TFA-MT in selected cases.
KEYWORD
Stroke, Radial, Thrombectomy, Guide catheter, Sheath, Aspiration
FullTexts / Linksout information
Listed journal information