Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1812320220390020108
Journal of Yeungnam Medical Science
2022 Volume.39 No. 2 p.108 ~ p.115
Clinical performance of FractionLab in patient-specific quality assurance for intensity-modulated radiotherapy: a retrospective study
Oh Se-An

Kim Sung-Yeop
Park Jae-Hyeon
Park Jae-Won
Yea Ji-Woon
Abstract
Background: This study was aimed at comparing and analyzing the results of FractionLab (Varian/Mobius Medical System) with those of portal dosimetry that uses an electronic portal imaging device. Portal dosimetry is extensively used for patient-specific quality assurance (QA) in intensity-modulated radiotherapy (IMRT).

Methods: The study includes 29 patients who underwent IMRT on a Novalis-Tx linear accelerator (Varian Medical System and BrainLAB) between June 2019 and March 2021. We analyzed the multileaf collimator DynaLog files generated after portal dosimetry to evaluate the same condition using FractionLab. The results of the recently launched FractionLab at various gamma indices (0.1%/0.1 mm?1%/1 mm) are analyzed and compared with those of portal dosimetry (3%/3 mm).

Results: The average gamma passing rates of portal dosimetry (3%/3 mm) and FractionLab are 98.1% (95.5%?100%) and 97.5% (92.3%?99.7%) at 0.6%/0.6 mm, respectively. The results of portal dosimetry (3%/3 mm) are statistically comparable with the QA results of FractionLab (0.6%/0.6 mm?0.9%/0.9 mm).

Conclusion: This paper presents the clinical performance of FractionLab by the comparison of the QA results of FractionLab using portal dosimetry with various gamma indexes when performing patient-specific QA in IMRT treatment. Further, the appropriate gamma index when performing patient-specific QA with FractionLab is provided.
KEYWORD
Electronic portal imaging device, FractionLab, Gamma passing rate, Intensity-modulated radiotherapy, Patient-specific quality assurance
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed