Chiral Purity Test of Bevantolol by Capillaryelectrophoresis and High Performance Liquid Chromatography
|
|
Long Pham Hai
Trung Tran Quoc Oh Joung-Won Kim Kyeong-Ho
|
|
Abstract
|
|
|
Two methods for the chiral purity determination of bevantolol were developed, namely capillary electrophoresis (CE) using carboxymethyl-¥â-cyclodextrin (CM-¥â-CD) as a chiral selector and high-perfomance liquid chromatography (HPLC) using a chiral stationary phase. In the HPLC method, the separation of bevantolol enantiomers was performed on a Chiralpak AD-H column by isocratic elution with n-hexane-ethanol-diethylamine (10:90:0.1, v/v/v) as mobile phase. In the CE method, bevantolol enantiomers were separated on an uncoated fused silica capillary with 50 mM amonium phosphate dibasic adjusted to a pH 6.5 with phosphoric acid containing 15 mM CM-¥â-CD as running buffer. Validation data such as linearity, recovery, detection limit, and precision of the two methods are presented. The detection limits of S-(-)- bevantolol were 0.1% and 0.05% for CE and HPLC method, respectively and R-(+)-bevantolol were 0.15% and 0.05% for CE and HPLC method, respectively. There was generally good agreement between the HPLC and CE results.
|
|
KEYWORD
|
|
Bevantolol, HPLC, CE, Chiral separation, Optical purity
|
|