Archives of Pharmacal Research
2016 Volume.39 No. 2 p.213 ~ p.220
Development and validation of a highly sensitive LC?MS/MS method for the determination of acacetin in human plasma and its application to a protein binding study
Kim Sang-Bum

Lee Tae-Hun
Lee Hun-Seok
Song Chung-Kil
Cho Hyun-Jong
Kim Dae-Duk
Maeng Han-Joo
Yoon In-Soo
Abstract
A highly sensitive bioanalytical method for the quantification of acacetin in human plasma was developed and comprehensively validated using liquid chromatography-tandem mass spectrometry (LC?MS/MS). A minimal volume of human plasma sample (20 ¥ìL) was prepared by simple deproteinization with 80 ¥ìL of acetonitrile. Chromatographic separation was performed using Kinetex C18 column with an isocratic mobile phase consisting of water and acetonitrile (20:80, v/v) containing 0.1 % formic acid at a flow rate of 0.3 mL/min over a total run time of 2.0 min. Mass spectrometric detection was performed using multiple reaction-monitoring modes at the mass/charge transitions m/z 285.22 ¡æ 242.17 for acacetin and m/z 277.59 ¡æ 175.04 for chlorpropamide (internal standard). The calibration curve was linear over the range of 0.1?500 ng/mL with a lower limit of quantitation of 0.1 ng/mL. The coefficients of variation for both intra- and inter-day validation were less than 11.9 %, and the intra- and inter-day accuracy ranged from 96.8 to 108 %. Mean recovery of acacetin in human plasma was within the range of 91.5?95.6 %. This validated LC?MS/MS method was successfully applied to a human plasma protein binding study that indicated extensive and concentration-independent protein binding of acacetin in human plasma.
KEYWORD
Acacetin, LC?MS/MS, Human plasma, Protein binding
FullTexts / Linksout information
Listed journal information
SCI(E) MEDLINE ÇмúÁøÈïÀç´Ü(KCI)