End Tidal CO2 Monitoring with Salter Divided Nasal Cannula in Post-extubated Spontaneous Breating Patients
Àå¹®¼®
ÀÓÇýÀÚ/Á¶Çå/°ø¸íÈÆ/±è³­¼÷/À强ȣ
Abstract
For the evaluation of the correlation between the ETCO2 and the PaCO2 in post-extubated spontaneously breathing patients, we tested 30 patients in recovery room with Salter Divided Nasal Cannula, which have permanent barrier in the face piece and
a
dual
tubing set that allows end tidal sampling from one nare and oxygen delivery to the other.
When the wave form of capnogram looked regular and mormal, the value of ETCO2 was recorded and the arterial blood was taken from the radial artery to analize blood gas immediately.
@ES the results were a following:
@EN 1. The mean value of PaCO2 was 42.0 ¡¾ 4.8 mmHg.
2. The mean value of PETCO2 was 39.3 ¡¾ 5.1 mmHg.
3. The value of PETCO2 acquired with Salter Divided Nasal Cannula hase close positive correlation with the values of the PaCO2.
(PaCO2=0.75 X PETCO2+12.64, r=0.79, p<0.001)