Breakage of a thoracic epidural catheter during its removal
Lee Yong-Ho

Hwang Hee-Youn
Sim Woo-Seok
Yang Mi-Kyung
Lee Chul-Joong
Abstract
The breakage of an epidural catheter is a rare complication during the removal of a thoracic epidural catheter. There are many causes to breakage of an epidural catheter, such as the characteristics of the catheter itself, patient¡¯s factors (anatomy, position during insertion and removal of the catheter, and the BMI), and the difficulty of the procedure. Surgical removal is considered if there is a possibility that the retained catheter might cause neurological problems. We experienced a breakage of an epidural catheter during its removal, which led to surgical intervention. To prevent the catheter breakage during its removal, the catheter should be withdrawn without excessive tension. In addition, an understanding of the patient¡¯s anatomy is essential.
KEYWORD
Breakage, Catheter, Complications, Epidural, Thoracic