KMID : 0356920100590040270
|
|
Korean Journal of Anesthesiology 2010 Volume.59 No. 4 p.270 ~ p.274
|
|
Interruption in the supply of breathing gas during general anesthesia due to malposition of the vaporizer -A case report-
|
|
Kim Hyo-Jin
Kim Mi-Woon
|
|
Abstract
|
|
|
We report a case of interruption in the supply of breathing gas during general anesthesia caused by malposition of the Drager Vapor 2000¢ç vaporizer, which was accidentally tilted and lifted off the Selectatec manifold of the anesthesia machine. Because the patient was an 1-month-old infant, we couldn¡¯t check if he had experienced awareness with recall. We emphasize the importance of checking the anesthetic vaporizer after mounting it on the back bar of the anesthesia machine.
|
|
KEYWORD
|
|
Gas leak, Selectatec, Vaporizer
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|