Korean Journal of Anesthesiology
2010 Volume.59 No. 4 p.270 ~ p.274
Interruption in the supply of breathing gas during general anesthesia due to malposition of the vaporizer -A case report-
Kim Hyo-Jin

Kim Mi-Woon
Abstract
We report a case of interruption in the supply of breathing gas during general anesthesia caused by malposition of the Drager Vapor 2000¢ç vaporizer, which was accidentally tilted and lifted off the Selectatec manifold of the anesthesia machine. Because the patient was an 1-month-old infant, we couldn¡¯t check if he had experienced awareness with recall. We emphasize the importance of checking the anesthetic vaporizer after mounting it on the back bar of the anesthesia machine.
KEYWORD
Gas leak, Selectatec, Vaporizer