KMID : 0356920110610010024
Korean Journal of Anesthesiology
2011 Volume.61 No. 1 p.24 ~ p.29
The effect of tracheal tube size on air leak around the cuffs
Hwang Jin-Young

Park Sang-Hyun
Han Sung-Hee
Park Seong-Joo
Park Soo-Kyung
Kim Jin-Hee
Abstract
Background: This randomized single-blinded, cross-over study was done to evaluate the influence of the size of tracheal tubes on air leaks around the cuffs.

Methods: In a benchtop model, the number of longitudinal folds on the cuffs was evaluated for different sizes of tracheal tubes. In an anesthetized patient study, thirty patients scheduled for elective surgery under general anesthesia were included. After induction of anesthesia, the trachea was intubated with two sizes of tracheal tubes in a random sequence: in men, internal diameter of 7.5 mm and 8.0 mm; in women, internal diameter of 7.0 mm and 7.5 mm. After tracheal intubation with each tube, air leak pressures were evaluated at intracuff pressures of 20, 25 and 30 cmH2O by auscultation. To calculate the tracheal tube resistance (R), an inspiratory pause of 20% was applied and the resulting peak airway pressure (Ppeak), plateau pressure (Ppl) and mean expiratory tidal volume (Flow) were inserted in the formula R = (Ppeak - Ppl)/Flow.

Results: More longitudinal folds of the tracheal tube cuffs occurred in larger sized tubes compared to the smaller ones in a benchtop model. Air leakage was significantly less for the smaller tracheal tubes than for the larger ones for each gender at intracuff pressures of 20, 25 and 30 cmH2O. Tracheal tube resistances were not significantly altered by the size of tracheal tube.

Conclusions: The use of a smaller tracheal tube within an acceptable size can reduce air leakage around the cuff without significantly changing the tracheal tube resistance.
KEYWORD
Anesthesia, Intratracheal, Intubation